Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-17
4
DELAY
This menu allows the users to set an audible saturation delay. The
delay can be set to either 0, 5,10 or 15 seconds. The delay setting only
affects saturation alarms indications.
NOTE: In “Custom” mode (see Section 4, Operation) the instrument
will retain the Alarm Delay setting after a power cycle. In “Neo” or
“Adult” mode (see Section 4, Operation) the instrument will reset the
Alarm Delay to the hospital specifi ed setting after a power cycle.
MODE
The Radical-7 stores three types of modes: Adult, Neonatal or Custom
limits. Adult and Neonatal must be initially set and enabled (via
password protected screen, see Section 4, Password Operation) before
they can be selected.
ADULT
Any changes to settings on the Alarm menu will be reset to pre-defi ned
Adult defaults after a power cycle.†
CUSTOM
Any changes to settings on the Alarm menu will be retained after a
power cycle.*
NEO
Any changes to settings on the Alarm menu will be reset to pre-defi ned
Neonatal defaults after a power cycle.†
TYPES
SpO
2
(High)
SpO
2
(Low)
PULSE RATE
(High)
PULSE RATE
(Low)
ADULT
LIMITS†
Off 90% 140 BPM 50 BPM
NEONATAL
LIMITS†
100% 90% 180 BPM 100 BPM
CUSTOM
LIMITS*
Off* 90%* 140 BPM* 50 BPM*
NOTE: Limits are set at the factory to the values listed in this table.
*Once Custom values are changed by the user, they will be retained
after a power cycle.
†Adult and Neo settings can be changed (via a password protected
screen, see Section 4, Password Operation) to specifi c hospital/instrument
requirements. If the settings are changed, then any values changed by the
user will be returned to the instrument's default values after it is powered down.
Operation
Alarms (continued)