Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-3
4
Operation
averaging times delay the response of the oximeter and reduce the measured variations
of SpO
2
and pulse rate.
MASIMO SENSORS
Before use, carefully read the Masimo sensor Directions for Use.
Use only Masimo sensors for pulse oximetry or pulse CO-Oximetry measurements.
Tissue damage can be caused by incorrect application or use of a sensor, for example by
wrapping the sensor too tightly. Inspect the sensor site as directed in the sensor Directions for
Use to ensure skin integrity and correct positioning and adhesion of the sensor.
NOTE: When a Masimo Rainbow Sensor is properly connected to the patient, the instrument
normally goes through a 20-30 second sensor calibration/pulse search routine and then displays
numeric values for SpO
2
, SpMet, SpCO, SpHb, SpOC, pulse rate, Perfusion Index (PI) and
Pleth Variability (PVI). It also provides graphical displays for plethysmographic waveform, Signal
Identification and Quality Indicator (Signal IQ
®
). However, if the sensor calibration/pulse search
routine is unsuccessful for Rainbow parameters/measurements, the instrument automatically
switches to a "SpO2 Only Mode" to provide SpO
2
, PR, PI and PVI parameters/measurements
for the user.
If a Masimo Rainbow Direct Connect Reusable Sensor is being used and "SpO2 Only Mode"
appears on the display screen, perform one of the following steps to reset the instrument:
Remove the sensor from from patient (recommended).
Remove the cable connector from instrument.
Turn the power Off and On at the instrument.
If a Masimo Rainbow Adhesive Sensor is being used and "SpO2 Only Mode" appears on the
display screen, perform one of the following steps to reset the instrument:
Disconnect sensor cable connector from the patient cable connector
(recommended).
Remove patient cable connector from instrument.
Turn the power Off and On at the instrument.
Remove the sensor from the patient.
To obtain successful monitoring, refer to Section 4, Successful Monitoring.
CAUTIONS
DO NOT USE DAMAGED SENSORS. DO NOT USE A SENSOR WITH EXPOSED
OPTICAL OR ELECTRICAL COMPONENTS. DO NOT IMMERSE THE SENSOR
IN WATER, SOLVENTS, OR CLEANING SOLUTIONS (THE SENSORS AND
CONNECTORS ARE NOT WATERPROOF). DO NOT STERILIZE BY IRRADIATION,
STEAM, AUTOCLAVE OR ETHYLENE OXIDE UNLESS OTHERWISE INDICATED IN
THE SENSOR DIRECTIONS FOR USE. SEE THE CLEANING INSTRUCTIONS IN THE
DIRECTIONS FOR USE FOR ALL MASIMO REUSABLE SENSORS.
DO NOT USE DAMAGED PATIENT CABLES. DO NOT IMMERSE THE PATIENT
CABLES IN WATER, SOLVENTS, OR CLEANING SOLUTIONS (THE PATIENT CABLE
CONNECTORS ARE NOT WATERPROOF). DO NOT STERILIZE BY IRRADIATION,
STEAM, AUTOCLAVE OR ETHYLENE OXIDE.