3-4 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
3
OPTION
FACTORY DEFAULT
SETTING
CONFIGURABLE
SETTING
TREND DISPLAY
PARAMETERS/
MEASUREMENTS
Set to %SpO
2
+ BPM
See Section 4, Display for
all settings.
TREND PERIOD
Set to 1 Hr
See Section 4, Trend
Setup for all settings.
SpHb TREND
Set to Yes Yes/No
LOW SpO
2
ALARM LIMIT
Set to 0
(Refer to Section 4,
Display for details on this
user adjustable feature)
SpHb AVERAGING
Set to Medium Short, Medium, Long
SpMet HIGH/LOW ALARM LIMIT
Set to High: 3.0/Low: ---
See Section 4, Alarms for
all settings.
SpCO HIGH/LOW ALARM LIMIT
Set to High: 10/Low: ---
SpHb HIGH/LOW ALARM LIMIT
Set to High: 17.0/Low: 7.0
PI
HIGH/LOW ALARM LIMIT
Set to High: ---/Low: ---
PVI
HIGH/LOW ALARM LIMIT
Set to High: ---/Low: ---
SMART TONE
Set to No Yes/No
PI AVERAGING
Set to Short Short, Long
PVI AVERAGING
Set to Long Short, Long
DISPLAY MEASUREMENT
SpOC SpOC, None
SpHb PRECISION
Set to 0.1 0, 0.1, 0.5
SpHb CAL
Arterial Arterial, Venous
ADULT/NEO SETTINGS (PRE-DEFINED)
The following table outlines settings that may be defined by the hospital. If enabled (via a password
protected screen, see Section 4, Password Operation), these settings will return to pre-defined values
after a power cycle. This mode is indicated by “Mode Adult” or “Mode Neo” on the Alarms menu. See
Section 4, Operation for details on enabling and setting these parameters/measurements.
OPTION
CUSTOM SETTING
(DEFAULT)
ADULT/NEO
SETTING (PRE-
DEFINED)
SPO
2
HIGH/LOW ALARM LIMIT
Set to High: ---/Low: 90
Same settings
PULSE RATE HIGH/LOW ALARM LIMIT
Set to High: 140/Low: 50 BPM
SpMet HIGH/LOW ALARM LIMIT
Set to High: 3.0/Low: ---
SpCO HIGH/LOW ALARM LIMIT
Set to High: 10/Low: ---
SpHb HIGH/LOW ALARM LIMIT
Set to High: 17.0/Low: 7.0
PI
HIGH/LOW ALARM LIMIT
Set to High: ---/Low: ---
PVI
HIGH/LOW ALARM LIMIT
Set to High: ---/Low: ---
ALARM SILENCE
Set to 120 seconds
ALARM VOLUME
Set to 3
ALARM DELAY
Set to 5
SpHb CAL
Set to Arterial
Setup
CUSTOM (USER) DEFINED SETTINGS (CONTINUED)