Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-15
4
MENU ITEMS DESCRIPTION
SpCO LOW LIMIT
The SpCO low alarm limit can be set as “---”, or anywhere between 1%
and 97%, with a 1% step size. In the “----” (off) setting, the alarm can be
turned off completely.
NOTE: The low alarm limit must be set below the high alarm setting. If
the high alarm limit is set below the low alarm limit, the low alarm limit
will automatically adjust to the next setting below the newly entered
high alarm limit setting.
SpHb HIGH LIMIT
The SpHb high alarm limit can be set anywhere between 2.0 g/dl and 24.5
g/dl, then “---” with a 0.1 g/dl step size between 2.0 and 20.0, and a 0.5 g/
dl step size between 20.0 and 24.5. In the “---” (off) setting, the SpHb High
Alarm Limit Alarm is disabled. Factory default setting is 17 g/dl.
SpHb LOW LIMIT
The SpHb low alarm limit can be set as “---”, or anywhere between 1.0
g/dl and 24 g/dl with a 0.1 g/dl step size between 1.0 and 20.0, and a
0.5 g/dl step size between 20.0 and 24.0. In the “---” (off) setting, the
SpHb Low Alarm Limit Alarm is disabled. Factory default setting is 7 g/dl.
NOTE: The low alarm limit must be set below the high alarm setting. If
the high alarm limit is set below the low alarm limit, the low alarm limit
will automatically adjust to the next setting below the newly entered
high alarm limit setting.
PI HIGH LIMIT
The PI high alarm limit can be set anywhere between 0.04 and 19,
then “---” with a 0.01 step size between 0.04 and 0.10, a 0.10 step size
between 0.10 and 1.0, and a 1.0 step size between 1.0 and 19. In the
“---” (off) setting, the PI High Alarm Limit Alarm is disabled. Factory
default setting is “---” (off).
PI LOW LIMIT
The PI low alarm limit can be set as “---”, or anywhere between 0.03
to 18 with a .01 step size between 0.03 and 0.10, a 0.10 step size
between 0.10 and 1.0, and a 1.0 step size between 1.0 and 18. In
the “---” (off) setting, the PI Low Alarm Limit Alarm is disabled. Factory
default setting is “---” (off).
NOTE: The low alarm limit must be set below the high alarm setting. If
the high alarm limit is set below the low alarm limit, the low alarm limit
will automatically adjust to the next setting below the newly entered
high alarm limit setting.
PVI HIGH LIMIT
The PVI high alarm limit can be set anywhere between 2 and 99, then “---”
with a 1 step size between 2 and 99. In the “---” (off) setting, the PVI High
Alarm Limit Alarm is disabled. Factory default setting is “---” (off).
PVI LOW LIMIT
The PVI low alarm limit can be set as “---”, or anywhere between 1 and
98 with a 1 step size. In the “---” (off) setting, the PVI Low Alarm Limit
Alarm is disabled. Factory default setting is “---” (off).
NOTE: The low alarm limit must be set below the high alarm setting. If
the high alarm limit is set below the low alarm limit, the low alarm limit
will automatically adjust to the next setting below the newly entered
high alarm limit setting.