Service and repair
REPAIR POLICY
Masimo or an authorized Service Department must perform warranty repair and service.
Do not use malfunctioning equipment. Have the unit repaired.
WARNING: DO NOT REMOVE THE COVER OF THE MONITOR EXCEPT FOR
BATTERY REPLACEMENT. AN OPERATOR MAY ONLY PERFORM
MAINTENANCE PROCEDURES SPECIFICALLY DESCRIBED IN
THIS MANUAL. REFER SERVICING TO QUALIFIED SERVICE PER-
SONNEL TRAINED IN THE REPAIR OF THIS EQUIPMENT.
Please clean contaminated and/or dirty equipment before returning, following the cleaning
procedure described in Section 9, Cleaning. Make sure it is fully dry before packing the
equipment.
To return the Radical-7 unit for service, please follow the Return Procedure.
RETURN PROCEDURE
Please clean contaminated/dirty equipment before returning and make sure it is fully
dry before packing the equipment. Call Masimo at 800-326-4890 and ask for Technical
Support. Ask for an RMA number. Package the equipment securely – in the original ship-
ping container if possible – and enclose or include the following information and items:
A letter describing in detail any difficulties experienced with the Pulse CO-Oximeter.
Please include the RMA number in the letter.
Warranty information – a copy of the invoice or other applicable documentation
must be included.
Purchase order number to cover repair if the oximeter is not under warranty, or for
tracking purposes if it is.
Ship-to and bill-to information.
Person (name, telephone/Telex/fax number, and country) to contact for any ques-
tions about the repairs.
A certificate stating the oximeter has been decontaminated for bloodborne pathogens.
Return Radical-7 pulse oximeter to the following shipping address:
For USA and Asia Pacific: For Europe:
Masimo Corporation
40 Parker
Irvine, California 92618
949-297-7000
FAX 949-297-7001
Masimo Europe Limited
304 RN6, Le Bois des Cotes 2
69760 Limonest
France
Analog Output Test
1. Disconnect all patient cables and sensors from the Radical-7. Turn the Radical-7 off
and then on again.
2. Connect the common lead of a digital voltmeter to the pin 2 (Ground) of the analog
output connector on the Radical-7. Connect the positive lead of the voltmeter to pin
9 (Analog 1) of the analog output connector.
3. Enter the menu system and set the “Output”, “Analog 1” to “0V Signal”. Verify that the
voltmeter measures a voltage of approximately 0V.
4. Enter the menu system and set the “Output”, “Analog 1” to “1V Signal”. Verify that the
voltmeter measures a voltage of approximately 1.0V.
5. Repeat Steps 3 and 4, with the positive lead of the voltmeter connected to pin 15
(Analog 2).
6. Connect a patient cable and sensor and verify that the voltage on pins 9 and 15 are
between 0V and 1.0V while measuring a saturation and pulse rate..
Battery Test
1. Fully charge the Radical-7 by placing the Handheld into the Docking Station and
connecting the AC power.
2. Verify that the green Handheld Battery Indicator LED is lit up.
3. When the Radical-7 is fully charged the green Handheld Battery Indicator turns off.
4. Turn the Radical-7 on and verify that the Battery indicator shows a full charge.