Masimo Radical Signal Extraction Instrukcja Operatora Strona 34

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 33
4-26 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 4-27
operation
4
4
operation
TREND SETUP
This menu allows the user to set the default trend settings or download the trend data t. The
default settings are used to scale the trend graphs when the trend data button, located on
the main display, is accessed.
MENU ITEMS DESCRIPTION
%SpO
2
MAX SCALE
Sets the maximum scale of the SpO
2
trend graph.
%SpO
2
MIN SCALE
Sets the minimum scale of the SpO
2
trend graph.
PR MAX SCALE
(BPM)
Sets the maximum scale of the pulse rate trend graph.
PR MIN SCALE
(BPM)
Sets the minimum scale of the pulse rate trend graph.
PI MAX SCALE
Sets the maximum scale of the PI trend graph.
PI MIN SCALE
Sets the minimum scale of the PI trend graph.
%SpMET MAX
SCALE
Sets the maximum scale of the SpMet trend graph.
%SpMET MIN
SCALE
Sets the minimum scale of the SpMet trend graph.
%SpCO MAX
SCALE
Sets the maximum scale of the SpCO trend graph.
%SpCO MIN SCALE
Sets the minimum scale of the SpCO trend graph.
DEFAULT VIEW
Selects the default time scale of the trend view. This setting only selects
the time scale of the trend view when the trend data is initially displayed,
(i.e. when the trend data is initially accessed). The selections are 24 hrs,
12 hrs, 8 hrs, 4 hrs, 2 hrs, 1 hr, 30 minutes, 10 minutes, 1 minute and
20 seconds.
MENU ITEMS DESCRIPTION
TREND ACTION
SERIAL DUMP
To send all the data that is stored in trend memory to the serial port se-
lect the Serial Dump option. Use this option to communicate the stored
data set to trend graphing software applications.
ANALOG DUMP
To send all the data that is stored in the trend memory to the analog
output select the Analog Dump option. Use this option to print the trend
information on an analog chart recorder.
PRINT
To print the trend data that is shown in the Trend View select the Print
option. The trend data is fi rst printed in histogram format, followed by a
table of data that shows the time and date stamp of a trend record, and
the SpO
2
, pulse rate, SpMet, SpCO and PI measurement. Each trend
record is printed on a single line followed by a carriage return and line
feed character.
TREND PERIOD
The Trend period setting determines how often a set of SpO
2
, pulse rate,
SpMet, SpCO and PI data points is stored in trend memory. A setting
of 2, for example, sets the Radical-7 Pulse CO-Oximeter to store one
set of SpO
2
, pulse rate, SpMet, SpCO and PI measurements every 2
seconds, resulting in a minimum trend capacity of 72 hours. A setting
of 10, for example sets the Radical-7 Pulse CO-Oximeter to store one
set of data points every 10 seconds, resulting in a typical trend storage
capacity of 18 days.
NOTE: Since the Radical-7 Pulse CO-Oximeter employs a sophisticated data compression
scheme, the actual trend capacity is dependent on the type of data that is collected.
HISTOGRAM
The figure below is an example of the display output of the Histogram function. The
Histogram displays the trend data of selected parameters as numerical percentages shown
in specific ranges.
Signal Extraction Pulse CO-Oximeter
Masimo
rainbow
SET
99
85
Sp0
2
140
50
BPM
1 Hr Trend
11:23:11pm
12:23:11pm
10/11/06
SpMet
SpCO
SpO
2
Max
Min
100
50
15.0
1.0
40
1.0
PI
BPM
Max
Min
200
25
20
1
1 min
Trend Action
Trend Period
None
2
Default View
Signal Extraction Pulse CO-Oximeter
Masimo
rainbow
SET
99
85
Sp0
2
140
50
BPM
1 Hr Histogram
%SpO
2
0
0
100
0
0
0
0
96-100
91-95
86-90
81-85
1-800
PI
0
0
0
83
0
0
17
16.1-20.0
12.1-16.0
8.1-12.0
4.1-8.0
0.1-4.00
Przeglądanie stron 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 54 55

Komentarze do niniejszej Instrukcji

Brak uwag