4-20 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 4-21
operation
4
4
operation
Output
NOTE: The output menu selections are only available when the Radical-7 Handheld is
interfaced to the Docking Station.
MENU ITEMS DESCRIPTION
SERIAL
The following serial output modes are supported. All serial output is
RS-232 based. See the interface specifi cations in Section 7-Specifi ca-
tions.
ASCII 1
ASCII text data is sent to the serial interface at one-second intervals.
The ASCII text includes: date and time stamp, SpO
2
pulse rate, PI,
SpMet, SpCO and alarm and exception values. All text is single line
followed by a line feed character and a carriage return.
ASCII 2
ASCII text data is sent to the serial interface following a query from the
connecting computer. This mode will need to be active for RadNet data
output.
BINARY
Compressed binary data is sent to the serial interface following a query
from the connecting computer.
PHILIPS VUELINK
SpO
2
, pulse rate and plethysmographic waveform data are sent in
Philips VueLink format to the serial port.
SPACELABS FLEXPORT
SpO
2
, pulse rate and plethysmographic waveform data are sent in
Spacelabs Flexport format to the serial port.
ANALOG1 OR
ANALOG 2
SpO
2
0 - 100%
Scales the saturation measurement with 0% being equal to 0 Volt and
100% equal to 1 Volt.
SpO2 50 - 100%
Scales the saturation measurement with 50% being equal to 0 Volt and
100% equal to 1 Volt.
PULSE RATE
Scales the pulse rate measurement with 0 BPM being equal to 0 Volt,
and 250 BPM equal to 1 Volt.
PLETH
Traces the plethysmographic waveform as shown on the Radical-7
display.
SIGNAL IQ
Traces the Signal IQ waveform as shown on the Radical-7 display. A
full scale Signal IQ signal (100%) is represented as 1 Volt, while a zero
Signal IQ signal (0%) is represented as 0 Volt.
Output continued
MENU ITEMS DESCRIPTION
ANALOG1 OR
ANALOG 2
(CONTINUED)
0V OUTPUT
A 0 Volt calibration signal is mapped to the analog output. Use this sig-
nal for calibration of recording devices. (0 Volts represents a saturation
of 0% and a pulse rate of 0 bpm).
1V OUTPUT
A 1 Volt calibration signal is mapped to the analog output. Use this sig-
nal for calibration of recording devices. (1 Volt represents a saturation
of 100% and a pulse rate of 250 bpm).
NURSE CALL
ALARMS
The nurse call output will be activated based on alarm events.
LOW SIGNAL IQ
The nurse call output will be activated based on Low Signal IQ events.
ALARM & SIGNAL IQ
The nurse call output will be activated based on alarm and Low Signal
IQ events.
POLARITY
NORMAL
Standard polarity. See section 7, Analog output / nurse call specifi ca-
tions.
INVERT
This setting reverses the Normally Open and Normally closed contacts.
See section 7, Analog output / nurse call specifi cations.
CAUTION: TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY
EQUIPMENT TO THE SERIAL PORT ON THE BACK PANEL UNLESS THE
RADICAL-7 IS CONNECTED TO THE AC MAINS POWER SUPPLY.