Masimo Radical Signal Extraction Instrukcja Operatora Strona 28

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 27
4-14 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 4-154-15
4
operation
Rotate Screen
MENU ITEMS DESCRIPTION
LANDSCAPE 1
The viewing screen will be in a horizontal fashion when the Radical-7 is
in the upright, horizontal position.
VERTICAL 1
The viewing screen will be in a vertical fashion when the Radical-7 is in
the upright, vertical position.
LANDSCAPE 2
The viewing screen will be in an inverted horizontal fashion (rotated
180
o
) when the Radical-7 is in the upright, horizontal position.
VERTICAL 2
The viewing screen will be in an inverted vertical fashion (rotated 180o)
when the Radical-7 is in the upright, vertical position.
MENU ITEMS DESCRIPTION
VOLUME
This menu allows the user to set the alarm volume. Four levels are
available: level 1 being the softest and level 4 being the loudest. The
device retains the Alarm Volume setting upon a power cycle.
NOTE: For home use, set the alarm level to level 4.
DELAY
This menu allows the users to set an audible saturation delay. The
delay can be set to either 0, 5 or 10 seconds. The delay setting only
affects saturation alarms indications.
NOTE: In “Custom” mode (see Section 4-Operation) the unit will retain
the Alarm Delay setting after a power cycle. In “Neo” or “Adult” mode
(see Section 4-Operation) the unit will reset the Alarm Delay to the
hospital specifi ed setting after a power cycle.
MODE
The Radical-7 stores three types of modes: Adult, Neonatal or Custom
limits. Adult and Neonatal must be initially set and enabled (via
password protected screen, see Section 4 Password Operation) before
they can be selected.
CUSTOM
Any changes to settings on the Alarm menu will be retained after a
power cycle.*
ADULT
Any changes to settings on the Alarm menu will be reset to pre-de-
ned Adult defaults after a power cycle.†
NEO
Any changes to settings on the Alarm menu will be reset to pre-de-
ned Neonatal defaults after a power cycle.†
TYPES
SpO
2
(HIGH)
SpO
2
(LOW)
PULSE RATE
(HIGH)
PULSE RATE
(LOW)
ADULT
LIMITS†
Off 90% 140 BPM 50 BPM
NEONATAL
LIMITS†
100% 90% 180 BPM 100 BPM
CUSTOM
LIMITS*
Off* 90%* 140 BPM* 50 BPM*
NOTE: Limits are set at the factory to the values listed in this table.
*Once Custom values are changed by the user, they will be retained
after a power cycle.
†Adult and Neo settings can be changed (via a password protected
screen, see Section 4 under Password Operation) to specifi c hospi-
tal/unit requirements. If the settings are changed, then any values
changed by the user will be returned to the unit's default values after it
is powered down.
operation
4-14
4
Przeglądanie stron 27