3-6 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 3-7
Philips VueLink Setup
CAUTIONS:
AFTER CONNECTING THE VUELINK CABLE TO THE SERIAL PORT, SATSHARE
MONITORING WILL NOT BE SUPPORTED.
A RADICAL DOCKING STATION THAT IS VUELINK COMPATIBLE MUST BE UTI-
LIZED.
SpMet AND SpCO CANNOT BE DISPLAYED ON THE MULTIPARAMETER MONI-
TOR WITH VUELINK.
1. Select the Philips VueLink selection from the Output menu on the Radical-7 Pulse
CO-Oximeter. Refer to Section 4, Output.
2. Connect one end of the VueLink cable to the Serial Output connector on the back
of the Docking Station.
3. Connect the other end of the VueLink cable to the VueLink module and insert the
module into the Philips monitor rack.
4. The SpO
2
and pulse rate values will automatically appear on the Philips monitor.
5. In order for the pleth waveform to be displayed on the Philips monitor and for
the Philips monitor to indicate the alarm conditions measured by the Pulse
CO-Oximeter, the user must configure the Philips monitor. Refer to the Philips
Operator’s Manual for complete instructions.
6. The Radical-7 can be set up to audibly indicate all patient alarms while commu-
nicating with the Philips VueLink module. Use the Interface Alarms setting in the
General menu to enable and disable audible alarms on the Radical-7.
Spacelabs Universal Flexport Setup
CAUTIONS:
AFTER CONNECTING THE SPACELABS FLEXPORT CABLE TO THE SERIAL
PORT, SATSHARE MONITORING WILL NOT BE SUPPORTED.
SpMet AND SpCO CANNOT BE DISPLAYED ON THE MULTIPARAMETER MONI-
TOR WITH FLEXPORT.
1. Select the Spacelabs Flexport selection from the Output menu on the Radical-7.
2. Connect one end of the Spacelabs Flexport cable to the Serial Output connector
on the back of the Docking Station.
3. Connect the other end of the Spacelabs Flexport cable to the Spacelabs
Universal Flexport connector.
4. The SpO
2
and pulse rate values will automatically appear on the Spacelabs
screen.
5. In order for the pleth waveform to be displayed on the Spacelabs screen and
for the Spacelabs monitor to indicate the alarm conditions measured by the
Pulse CO-Oximeter, the user must configure the Spacelabs monitor. Refer to the
Spacelabs monitor Operator’s Manual for complete instructions.
6. The Radical-7 Pulse CO-Oximeter can be set up to audibly indicate all patient
alarms while communicating with the Spacelabs Flexport module. Use the
Interface Alarms setting in the General menu to enable and disable audible
alarms on the Radical-7 Pulse CO-Oximeter.
3
setup
response). The Radical-7 Pulse CO-Oximeter’s ideal waveform necessitates the
need for additional averaging by the monitor. If the multiparameter monitor’s aver-
aging time is not changed, the time to display physiological changes in saturation
on the monitor will be increased with SatShare. However, the delay can be mini-
mized by reducing the multiparameter monitor’s averaging time.
7. While in the SatShare mode, if there are any significant discrepancies between
the readings from the Radical-7 Pulse CO-Oximeter and those on the monitor dis-
playing the values obtained from SatShare, the values reported by the Radical-7
Pulse CO-Oximeter are to be considered the correct values.
8. To use the Radical-7 Pulse CO-Oximeter with SatShare while it is not connected
to AC power, set the Power Save parameter in the General menu to “No” and refer
to Section 4, Operation. Please note that if the Radical-7 Pulse CO-Oximeter is
used in this mode, the length of time the Radical-7 Pulse CO-Oximeter can oper-
ate on battery power will be significantly diminished.
9. Set the SatShare Numbers and the Interface Alarms parameters in the General
menu according to Customer preference. The Interface Alarms cannot be dis-
abled if the SpCO and/or SpMet parameters are present in the Radical-7.
10. If displaying the simulated waveform is not desirable, it is recommended to turn
off the pleth waveform display of the multiparameter patient monitor.
SATSHARE SIGNALS ARE IDEAL SIMULATED WAVEFORMS CORRESPOND-
ING TO THE CALCULATED SATURATION AND PULSE RATE VALUES AND DO
NOT CONTAIN ALL OF THE INFORMATION CONTAINED IN PHYSIOLOGICAL
WAVEFORMS. THE MULTIPARAMETER PATIENT MONITOR DECODES THESE
SIGNALS INTO SATURATION AND PULSE RATE VALUES.
DURING SATSHARE OPERATION, THE AUDIBLE ALARMS MAY BE MUTED
ON THE RADICAL-7 PULSE CO-OXIMETER. WHEN THE AUDIBLE ALARM IS
MUTED (INDICATED BY THE BELL WITH A SLASH THROUGH IT) ON THE
RADICAL-7 PULSE CO-OXIMETER, USE THE MULTIPARAMETER MONITOR
FOR AUDIBLE ALARM INDICATION.
THE AUDIBLE ALARMS CANNOT BE MUTED IF THE SpCO AND/OR SpMet
PARAMETERS ARE PRESENT IN THE RADICAL-7.
DURING SATSHARE OPERATION DO NOT USE THE PLETH WAVEFORM DIS-
PLAY ON THE MULTIPARAMETER MONITOR FOR DIAGNOSTIC PURPOSES.
INSTEAD, USE THE PLETH WAVEFORM DISPLAYED ON THE RADICAL-7
PULSE CO-OXIMETER SCREEN.
TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY
EQUIPMENT TO THE SATSHARE CONNECTOR UNLESS THE RADICAL-7
PULSE CO-OXIMETER IS CONNECTED TO THE AC MAINS POWER SUPPLY.
ONLY USE A SATSHARE CABLE THAT HAS A FERRITE BEAD INSTALLED.
SpMet AND SpCO CANNOT BE DISPLAYED ON THE MULTIPARAMETER MONI-
TOR WITH SATSHARE.
WARNING: EXTERNAL DEVICE CONNECTIONS TO THE SATSHARE PORT
MUST BE IEC-60601-1-1 COMPLIANT.
setup
3