2-4 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 2-5
system description
2
19
APOD
MAX
SENSITIVITY
The MAX or APOD™ sensitivity icon is shown on the
Radical-7 display to indicate if the Radical-7 is set to operate
in Normal sensitivity, Maximum sensitivity or Adaptive Probe
Off Detection mode. When in Normal mode, this area will
appear blank.
20
ALARM SILENCE
BUTTON
Press the Alarm Silence Button to temporarily silence patient
and low battery alarms. Press the Alarm Silence Button
when the SENSOR OFF message is flashing (i.e. the sen-
sor is removed from the patient) to acknowledge the end of
monitoring. In this state, all further alarms are suspended
until the Pulse CO-Oximeter starts measuring SpO
2
, SpMet,
SpCO, PI and pulse rate again.
NOTE: System failure alarms can be silenced by pressing
the Power/Standby or Alarm Silence Button. If the Power/
Standby Button does not silence the system fault alarm,
press the Alarm Silence Button.
21
FastSat
FASTS AT
The FastSat
®
label is shown on the Radical-7 display when-
ever the Radical-7 is set to operate in the FastSat mode.
22
POWER/
STANDBY
BUTTON
Press the Power/Standby Button to turn the instrument on.
Press, hold the button for more than 2 seconds and then
release the button to turn the instrument off.
23
PI
PERFUSION
INDEX
The Perfusion Index indicates numerically the percentage of
pulsatile signal to non-pulsatile signal (pulse strength)
24
PULSE RATE
ALARM LIMITS
DISPLAY
The Pulse Rate Alarm Limits Display shows the upper and
lower pulse rate alarm limits. When an alarm limit is reached
or exceeded, the pulse rate value and the violated limit
flashes. (may change)
25
PULSE RATE
The Pulse Rate Measurement Display shows the patient’s
pulse rate in beats per minute. The upper and lower pulse
rate alarm limits are also displayed next to the pulse rate
measurement. The pulse rate is calculated and the display is
updated at a frequency of once per second.
26
BRIGHTNESS
LEVEL
The Brightness level icon displays when utilizing the
Backlight/Contrast button.
27
5
PLETH
VARIABILITY
INDEX
The Pleth Variability Index indicates numerically the percent-
age of variation in the pleth waveform as a result of a inhala-
tion and exhalation cycle.
* For instruments that include SpMet and SpCO parameters: the SpMet and SpCO param-
eters will be grayed out in the display screen if a non-Rainbow sensor is being used.
NOTE: The Pulse CO-Oximeter display view will be different without the SpMet and SpCO
options.
system description
2-4
2
9
PULSE
WAVEFORM
DISPLAY
The Pulse Waveform Display shows the acquired plethysmo-
graph waveform. The pleth waveform is scaled with signal
strength. Signal strength is defined as the relation of arterial
pulsatile signal to the non-pulsatile signal component.
10
BATTERY
STATUS
INDICATORS
The Battery Status Indicators show the capacity of the
Radical-7 Handheld and optional Docking Station batteries.
The indicator flashes when less than 15 minutes of bat-
tery life is left and the battery needs to be recharged. The
Docking Station Battery Status indicator is not shown when
the optional Docking Station battery is not present.
11
TOUCH KEY
CONTROL
BUTTONS
Press a Touch Key Control Button to select the correspond-
ing touch key icon. See Section 4, Touch Key Control Buttons
and Icons for more details.
12
SPEAKER
The speaker indicates audio alarms. Care should be taken
not to cover the speaker and muffle the audible alarm
volume.
13
PATIENT CABLE
CONNECTOR
Connect the patient cable to the Handheld Radical-7 by
plugging the cable into the Patient Cable Connector. Use
only Masimo compatible sensors and cables with this
oximeter. See Section 8, Sensors and Patient Cables for
more details.
14
TOUCH KEY
ICONS
The Touch Key Icons indicate the software menu items that
can be selected through the Touch Key Control Buttons.
Pressing a Touch Key Control Button next to an icon selects
the option.
15
SIGNAL IQ
The Signal IQ shows the acquired signal quality and the
timing of the pulse. A tall vertical line indicates a high qual-
ity signal, while a small vertical line indicates a low quality
signal. The Signal IQ will be displayed as a single, pulsating
bar in the Numbers mode.
16
TIME AND DATE
INDICATOR
The Time and Date Indicator displays the current time and
date. The time is displayed in 12 or 24 hour format. The date
is displayed in dd/mm/yy or mm/dd/yy format. Select the date
and time display formats in the Clock menu.
Philips Vuelink
OUTPUT MODE
INDICATOR
The Output Mode Indicator displays the output mode select-
ed by the user (Philips Vuelink or Spacelabs Flexport). The
Output Mode Indicator also displays the type of SatShare
®
cable. The Output Mode Indicator is only displayed when the
Radical-7 device actively outputs data other than ASCII text,
or interfaces with a monitor through the SatShare cable. The
Output Mode Indicator will take place of the Time and Date
indicator when utilized.
HP01
17
RadNet
SERIAL
OUTPUT MODE
The Serial Output Mode displays the selected output interface
when interfaced with the Docking Station. and connected to the
serial port with the selected serial cable.
18
BACKLIGHT/
CONTRAST
BUTTON
Press the Backlight Button to change the illumination level of
the backlight. With the AC line power connected, four levels
of illumination are available (in addition to the no illumination
level). In the Handheld mode, three levels of illumination
are available (again in addition to the no illumination level).
Use the lowest illumination for most efficient battery usage.
The backlight Button is also used to change the contrast
of the LCD display. Press and hold the Backlight Button for
longer than six seconds to change the contrast. Release the
Backlight Button at the desired contrast setting.