
2-12 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 3-1
Introduction
Before the Radical-7 Pulse CO-Oximeter can be used in a clinical setting, it needs to be
inspected, properly setup and the batteries need to be fully charged.
Unpacking and inspection
Remove the instrument from the shipping carton and examine it for signs of shipping dam-
age. Check all materials against the packing list. Save all packing materials, invoice and bill
of lading. These may be required to process a claim with the carrier.
If anything is missing or damaged, contact the Technical Service Department. The contact
address and phone numbers are listed in Section 9, Service and Repair.
Preparation for monitoring
The following sections of the manual describe the preparation, set-up and initial installa-
tion of the Radical-7 Pulse CO-Oximeter.
RADICAL-7 DOCKING STATION POWER REQUIREMENTS
Always use a hospital grade, AC power cable to connect the Radical-7 Pulse CO-
Oximeter to an AC power source. Do not connect the Radical-7 Docking Station to an AC
outlet controlled by a switch because the power to the instrument may be inadvertently
be switched off.
Verify the AC power voltage and line frequency before use. Verify that the power source
can provide adequate power rating as indicated on the rear panel of the Radical-7 Docking
Station.
The Radical-7 Pulse CO-Oximeter is designed to operate on 100 to 240VAC, 47-63 Hz.
The device is rated at 55 VA max.
Connect a hospital grade power cable to the power entry module of the Radical-7 unit
(IEC-320 connector type at the unit). Connect the power cable to an AC power source.
Ensure that the unit is adequately powered by verifying that the AC power indicator on
the Docking Station is illuminated.
CAUTION:
■ DO NOT UNDER ANY CIRCUMSTANCES REMOVE THE GROUNDING CONDUCTOR FROM
THE POWER PLUG.
■ DO NOT USE EXTENSION CORDS OR ADAPTERS OF ANY TYPE. THE POWER CORD AND
PLUG MUST BE INTACT AND UNDAMAGED.
■ USE THE POWER CORD AS THE MEANS TO DISCONNECT THE DEVICE FROM THE
MAINS POWER SUPPLY.
■ IF THERE IS ANY DOUBT ABOUT THE INTEGRITY OF THE PROTECTIVE EARTH CON-
DUCTOR ARRANGEMENT, OPERATE THE OXIMETER ON INTERNAL BATTERY POWER
UNTIL THE AC POWER SUPPLY PROTECTIVE CONDUCTOR IS FULLY FUNCTIONAL.
■ TO ENSURE PATIENT ELECTRICAL ISOLATION, CONNECT ONLY TO OTHER EQUIP-
MENT WITH ELECTRICALLY ISOLATED CIRCUITS.
■ DO NOT CONNECT TO AN ELECTRICAL OUTLET CONTROLLED BY A WALL SWITCH OR
DIMMER.
3-1
3
setup
system description
2
Radical-7 Monitor Interface
In addition to being a full-featured Handheld and Standalone Pulse CO-Oximeter, the
Radical-7 Pulse CO-Oximeter’s unique SatShare interface links the Radical-7 Pulse CO-
Oximeter to most existing multiparameter patient monitors through the pulse oximetry
patient cable or SpO
2
input connector.
■ Upgrades any approved and validated monitor to Masimo SET performance by using
the calculated SpO
2
and pulse rate determined by Radical-7 to simulate an ideal
waveform, which is sent to the validated multiparameter patient monitor.
■ Connects into the SpO
2
patient cable or SpO
2
input connector of the multiparameter
patient monitor.
Refer to Section 3, SatShare Setup and Section 4, SatShare Operation for additional details.
CAUTIONS:
■ THE WAVEFORM DISPLAYED ON THE MULTIPARAMETER PATIENT MONITOR
IS A SIMULATED SIGNAL (NON-NORMALIZED). REFER TO THE RADICAL-7
PULSE CO-OXIMETER DISPLAY FOR PATIENT WAVEFORM.
■ IF DISPLAYING THE SIMULATED WAVEFORM IS NOT DESIRABLE, IT IS REC-
OMMENDED TO TURN OFF THE PLETH WAVEFORM DISPLAY ON THE MULTI-
PARAMETER MONITOR
■ ONLY USE A SATSHARE CABLE THAT HAS A FERRITE BEAD INSTALLED.
■ SpMet AND SpCO CANNOT BE DISPLAYED ON THE MULTIPARAMETER MONI-
TOR WITH SATSHARE.
S
i
g
n
a
l
E
x
t
r
a
c
t
i
o
n
P
u
l
s
e
C
O
-
O
x
i
m
e
t
e
r
97
76
-
-
-
1
4
0
5
0
9
5
%
S
p
O
2
B
P
M
F
a
s
t
S
a
tA
P
O
D
10
/1
1/
0
6
1
2
:2
3
:1
1
p
m
1
.
65
20
5
%
S
p
C
O
3
.
0
5
0
0
.
3
%
S
p
M
e
t
3
.
00
P
I
T
R
E
N
D
G
R
A
P
H
M
A
X
A
P
O
D
1
2
0
Sensor Off
5
P
V
I
Komentarze do niniejszej Instrukcji