Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual 5-1
5
alarms and messages
Alarm Indication
An alarm condition is indicated by an audible alarm tone and visual alarm indicator. An Out-
of-limit parameter will flash on the Rad-5 only. “SEn OFF” and “nO SEn” will only generate
an alarm condition after a pulse has been found.
Alarm limits (Rad-5 only)
An audible alarm and a flashing alarm status indicator will occur when an alarm limit is met
or exceeded for greater than five seconds. Directions for alarm suspension are indicated
below. When a sensor is not connected to a patient, or when a sensor is not connected to
its cable, the display will read SEn OFF or NO SEn. An audible alarm will accompany the
display unless the oximeter has been set to Alarm Suspend Mode.
SETTING RANGE
SpO
2
High Limit
The SpO
2
high alarm limit can be set anywhere between 2% and
100%, with a 1% step size. In the “----” (off) setting, the SpO
2
High
Limit alarm is disabled.
SpO
2
Low Limit
The SpO2 low alarm limit can be set anywhere between 1% and
100%, with a 1% step size.
Note: The low alarm limit must always be set below the high alarm setting.
Attempting to set the high alarm limit below the low alarm limit, the low alarm
limit will automatically adjust the low limit to the next setting below the newly
entered high alarm limit setting.
Pulse Rate High
Limit (BPM)
The pulse rate high alarm limit can be set anywhere between
30 BPM and 240 BPM, with a 5 BPM step size.
Pulse Rate Low
Limit (BPM)
The pulse rate low alarm limit can be set anywhere between
25 BPM and 235 BPM, with a 5 BPM step size.
Note: The low alarm limit must always be set below the high alarm setting.
Attempting to set the high alarm limit below the low alarm limit, the low alarm
limit will automatically adjust the low limit to the next setting below the newly
entered high alarm limit setting.
WARNING: TO ENSURE THAT ALARM LIMITS ARE APPROPRIATE FOR THE
PATIENT BEING MONITORED, CHECK THE LIMITS EACH TIME THE PULSE
OXIMETER IS USED.
4-16 Rad-5/5v Signal Extraction Pulse Oximeter Operator’s Manual
4
Special menu
This section gives an overview of the Rad-5 special menu selections available. To navigate
through the menus, use the
Mode/Enter
,
Next
,
Up
and
Down
keys located on the front
panel of the oximeter. The following sub-sections describe each menu item in more detail.
The oximeter has options that allow user configuration to suit specific needs. These selec-
tions apply only to the Rad-5
STANDARD AND SLEEP MODE
Turn instrument on, then push and hold the Mode/Enter and Next buttons simultaneously
for 3 seconds to enter the special menu level.
SLEEP MODE OPERATION
The Rad-5 can be placed into the Sleep Mode to allow the unit to capture normal and
abnormal data without the triggering alarms. This mode will blank out the unit display with
the exception of the Battery Level Indicator and the Alarm Silenced Indicator and disable
the alarms even after a power cycle. The keypad will be locked except for the power but-
ton. However, any single key press will bring the display back for 10 seconds. Upon power
up, the SLP mode will be displayed along with a 10 second display of parameters. The
Mode Enter and Next key held simultaneously for 3 seconds (select next (STD), Mode
Enter) will put it back into the special menu to exit.
CAUTION: ALARMS ARE DISABLED IN THIS MODE
BUTTONS SETTINGS
+
Simultaneous
for 3 seconds
Standard Mode (STD)
Use
Up
or
Down
Arrow
Keys to adjust
parameter to
desired setting.
Note: The
parameter is
set/selected
when Mode
Enter or Next
are pressed.
Sleep Mode
(SLP)
LED Display Brightness
(4 levels) Note: Only
availabile Indicators are
illuminated while adjust-
ing setting.