Masimo Radical Signal Extraction Instrukcja Operatora Strona 60

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 59
Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 7-1
7
specifications
Rad-8 specifications
PERFORMANCE
Measurement Range
Arterial Oxygen Saturation (%SpO
2
): 1-100%
Pulse Rate: 25-240 beats per minute (bpm)
Perfusion Index 0.02% - 20%
Response time: <1 second delay
ACCURACY
Arterial Oxygen Saturation Accuracy
1
Saturation 60% to 80%
No Motion
2
Adults, Infants, Pediatrics ±4%
Saturation 70% to 100%
No Motion
3
Adults, Infants, Pediatrics ± 2%
Neonates ± 3%
Motion
4
Adults, Infants, Pediatrics, Neonates
± 3%
Low Perfusion
5
Adults, Infants, Pediatrics, Neonates ± 2%
Pulse Rate Accuracy
6
Pulse Rate:
25 - 240 (bpm)
No Motion
Adults, Infants, Pediatrics, Neonates
± 3 bpm
Motion
5
Adults, Infants, Pediatrics, Neonates
± 5 bpm
Low Perfusion
Adults, Infants, Pediatrics, Neonates
± 3 bpm
Resolution
Arterial Oxygen Saturation (%SpO
2
)
1%
Pulse Rate
1 bpm
ELECTRICAL
AC Power requirements: 100-240 VAC, 47-63 Hz
Power consumption:
20 VA max.
Battery
Type: Sealed lead acid
Capacity: (battery life) up to 7 hours
7
Charging time: 8 hours
ENVIRONMENTAL
Operating Temperature: 41°F to 104°F (5°C to 40°C)
Transportation/Storage Temperature: -40°F to 158°F (-40°C to +70°C)
8
Storage Humidity: 5% to 95%, non-condensing
Operating Altitude:
500 mbar to 1060 mbar pressure,
-1000 ft to 18,000 ft (-304 m to 5,486 m)
Przeglądanie stron 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 78 79

Komentarze do niniejszej Instrukcji

Brak uwag